User`s guide

Imaging Modalities
2D Imaging
Revision D.0 7-29
ssn February 10, 1999 C:\WINNT\Profiles\dapowell\Desktop\D.0 Books\CD FILES SONOS
D.0\System Basics D.0\Frame Files\7CH.FM add.2
Setup
Acquire
Max:120/300Hz
Works in conjunction with the Maximize Transfer/
Frames touch control in Disk Setup. Acquire
Max:120 Hz limits frame rate to 120 Hz. Acquire
Max:300 Hz allows frame rates up to 300 Hz.
Depth Marks Displays fixed depth marks along the right side of the
image. The MMode/Doppler line is displayed only if
MMode or Doppler is on while depth marks are
displayed.
Time Gain
Compensation
(TGC) sliders
Adjusts amplification of returning 2D signals at a
specific image depth. Moving the slider to the right
increases amplification; moving the slider to the left
decreases amplification. Screen display of slider
settings is toggled on and off using the TGC Curve 2D
Setup control.