User`s guide
Imaging Modalities
2D Imaging
Revision D.0 7-27
ssn February 10, 1999 C:\WINNT\Profiles\dapowell\Desktop\D.0 Books\CD FILES SONOS 
D.0\System Basics D.0\Frame Files\7CH.FM add.2
TGC Curve
Turns the TGC Curve display on or off. Press   
and touch TGC Curve. Then press 
 again to 
return to the live display. TGC sliders affect the screen 
display only in 2D and Tissue Doppler modes. 
Trapezoid Activates trapezoidal imaging, which widens the linear 
image to a trapezoidal shape. Available only with 
transducers that offer this feature.
U/D Invert Inverts the apex of the sector from its current
orientation. Appears on the right primary touch panel in 
cardiac presets and on the secondary touch panel in 
noncardiac presets. Also appears on the left touch panel 
under the Image controls. Available only with sector 
and curvilinear array transducers.
Zoom
Turns zoom (magnification) on or off. To move the 
zoom preview area, use the trackball (  is active 
when you first enter zoom preview). To resize the zoom 
preview area, press   and use the trackball. Then 
touch Zoom again to enter full zoom mode.
Setup
Setup
Tip: When activated, 
Zoom is first outlined 
by a box. This 
indicates zoom 
preview. Touch Zoom 
a second time to go 
into full zoom. 
Position
Size