User`s guide
Imaging Modalities
Harmonic Fusion Imaging
7-12 Revision D.0
Controls
Primary
Secondary
Harmonic Fusion 
(Touch Control)
Activates Harmonic Fusion for imaging tissue. 
Activates Harmonics for a 2D image only. Available 
only with a harmonics-ready transducer, such as the 
s3, s4, s8, 11-3L, c3540, OmniPlane III, or x4.
Harmonic Fusion 
(Rotary Control)
Selects the Harmonic Fusion setting and is available 
only when Harmonic Fusion is on. Changes are 
reflected in the Harmonic Fusion icon and on the 
process line displayed at the top left corner of the 
screen. Not available when the image has been 
frozen using the 
 key.
Harmonics Plus Provides enhanced Harmonics when the s3 
transducer is in Harmonic Fusion settings 1, 3, and 4. 
May reduce the display frame rate.
L/R Invert Switches the left/right orientation of the image. 
Appears on the right primary touch panel in
noncardiac presets. Also appears on the left touch 
panel under the Image controls.
Persist Adjusts the amount of frame averaging, which can 
give a smoother appearance. 
Power Adjusts transmit power in decibels (dB) where
0.0dB = maximum power and -30.0dB = minimum 
power. 
Secondary
Controls
Switches between the primary and secondary touch 
panels. When the control is highlighted, the
secondary controls are active.
Freeze