User`s guide
Imaging Modalities
Harmonic Fusion Imaging
7-12 Revision D.0
Controls
Primary
Secondary
Harmonic Fusion
(Touch Control)
Activates Harmonic Fusion for imaging tissue.
Activates Harmonics for a 2D image only. Available
only with a harmonics-ready transducer, such as the
s3, s4, s8, 11-3L, c3540, OmniPlane III, or x4.
Harmonic Fusion
(Rotary Control)
Selects the Harmonic Fusion setting and is available
only when Harmonic Fusion is on. Changes are
reflected in the Harmonic Fusion icon and on the
process line displayed at the top left corner of the
screen. Not available when the image has been
frozen using the
key.
Harmonics Plus Provides enhanced Harmonics when the s3
transducer is in Harmonic Fusion settings 1, 3, and 4.
May reduce the display frame rate.
L/R Invert Switches the left/right orientation of the image.
Appears on the right primary touch panel in
noncardiac presets. Also appears on the left touch
panel under the Image controls.
Persist Adjusts the amount of frame averaging, which can
give a smoother appearance.
Power Adjusts transmit power in decibels (dB) where
0.0dB = maximum power and -30.0dB = minimum
power.
Secondary
Controls
Switches between the primary and secondary touch
panels. When the control is highlighted, the
secondary controls are active.
Freeze