User`s guide

Imaging Modalities
Harmonic Fusion Imaging
Revision D.0 7-11
ssn February 10, 1999 C:\WINNT\Profiles\dapowell\Desktop\D.0 Books\CD FILES SONOS
D.0\System Basics D.0\Frame Files\7CH.FM add.2
The Harmonic Fusion icon on the far left shows penetration selected, as indicated
by a circled P, and shows the transducer transmitting at 1.3 MHz and receiving at
2.6 MHz. The Harmonic Fusion icon on the far right shows resolution selected,
as indicated by a circled R, and shows the transducer transmitting at 1.6 MHz and
receiving at 3.2 MHz. The Harmonic Fusion setting in the center shows texture
selected, as indicated by a circled T, and shows the transducer transmitting at 1.3
MHz and receiving at 2.6 MHz. The other two intermediate Harmonic Fusion
settings show a mixture of penetration and texture, or texture and resolution. The
icons show circled P and T, or T and R.
NOTE The Harmonic Fusion rotary control is not displayed when in Tissue Doppler
mode.
For a complete list of transducers that are used in harmonic imaging, see the
Transducer Reference.