Owner's Manual
Tru Pointe
Instruction 0019-9328
4
Technical Characteristics
Power ..........................................Two ‘D’ Alkaline Batteries
Warm-Up Time .......................... 10 seconds
Refrigerants Detected ................ All CFC, HCFC and HFC refrigerants
including:R-12,R-22,R-123,R-134aand
blendsR-404A,R-408A,R-409A,and
R-410A
Sensor:
Type ......................................... Heated Diode, Dispersive Electron, Plug-In
LifeExpectancy* ..................... Typically150hoursor1yearofnormaluse
Battery Life ................................ 11 hours min. high sensitivity mode,
20 hours min. low sensitivity mode,
as tested using Duracell Ultra MX1300
alkaline batteries
Response Time ........................... Instantaneous
Sensitivity................................... The instrument will detect a leak of at
least0.5oz/year(14g/yr)asdetectedby
movingprobetipat2"(50mm)persec-
ond,0.2"(5mm)aboveleaksource.Aleak
rateoflessthan0.1oz/year(3g/yr)can
be detected when probe tip is held steady
overleaksourceforatleast5seconds.
Sensitivity Adjustment .............. Automatic
Weight ........................................ 1.16lb(0.53kg)w/batteries
Size(excludingprobe) ............... 2.12Wx9.5Lx2.4Hinches
(53.8x241.3x61.0mm)
Probe Length .............................. 14"(356mm)
Technical Characteristics
* Caution: Exposing the sensor to a steady stream of highly concentrated
refrigerant will severely reduce sensor life or damage the sensor.
Sensor life is directly proportional to the amount of refrigerant
that passes through the sensor.