Owner's Manual
Tru Pointe
Instruction 0019-9328
13
Sensor Replacement
Over time the sensor will become less responsive to refrigerant gas as it
nearstheendofitsusefullife.Asensor’saveragelifeexpectancyisap-
proximately150hoursor1yearofnormaluse,butwillvarydependingon
the amount of refrigerant that passes through the sensor.
Replacethesensorasfollows(seeFigure4):
Material Required:
•Sensor(refertoSectionReplacement Parts)
Procedure:
1. Turn OFF instrument.
2. Grasp nose cone of probe assembly and pull probe away from case
using a slight twisting motion.
3. Sensor may be hot! Pull spent sensor from its socket and discard.
4. Pluginnewsensor.
5. Pushprobeassemblybackintocase.
6. Turn ON the instrument and check that it responds in a positive man-
ner to a source of refrigerant gas.
Maintenance
Figure 4. Sensor Replacement
Grasp probe assembly’s nose
cone and pull probe away from
case while applying a slight
twisting motion.
REFRIGERANT
SENSOR