Specifications

SF
6
Portable Monitor
Instruction 3015-5150 27
4.2 Internal Particulate / Hydrophobic Filter Replacement
Items Required:
Replacement filter (P/N 0007-1654)
Medium Phillips head screwdriver
Procedure:
1. Disassemble the monitor’s metal chassis per Section 4.1.
2. See illustration below and locate the internal particulate / hydrophobic filter.
3. Pull off tubing from both ends of filter and remove filter from instrument.
4. Attach tubing to new filter
5. Reassemble monitor.