3.2.3 Connecting the Exhaust Line (optional)
The optional exhaust line carries the exhausted gas sample away from the monitor.
The exhaust line can be up to 50 ft (15.2 m) in length. Ideally this line should terminate outdoors in a
location that is not exposed to the elements.
Connect the exhaust line to the monitor by firmly pushing the tubing into the Exhaust Port’s push-to-
connect (PTC) fitting. (See previous figure.) If the exhaust line terminates outside the building, position
the tubing so that no water or moisture can enter the line. Note that this line does not require a
termination filter.
To connect the exhaust line to the exhaust
port’s PTC fitting, firmly push the appropriate
tubing into the hole in the center of the
connector until it seats in the connector. Refer
to the figures at the left.
To remove tubing from a PTC connector, push
and hold the spring-loaded collar inwards,
then simultaneously withdraw the tubing.
Refer to the figures at the right.