Specifications

Monoxor
Plus Manual Configuration
0019-9370 Rev 0 17
Ambient CO Function
The test results can be printed by pressing F1 and
saved to memory (with a time and date stamp) by
pressing F3. Press F2 to return to the menu.
NOTE: Any over-range CO values (e.g., CO
> 2000 ppm) are displayed as “xxx”.
3.5. Differential Temperature Menu
Diff Temp Function
The Diff Temp option is used to calculate a differential
temperature based on two sampled temperatures.
temperature (Temp) is displayed. Values for T1 and T2
will show no readings until they are saved by the
operator, after which a differential value (T1-T2) will
also be displayed.
With temperature #1 (T1) selected, p
thermocouple in the target location, and wait for the
temperature reading (Temp) to stabilize. Press the
SAVE button to temporarily store the T1 value. Repeat
these steps for the temperature #2 (T2). After saving
both values, the “T1-T2” differential temperature will
be displayed. At this point, you may press SAVE again
to save the differential temperature value to memory
so that it can be retrieved and/or printed later.
Note that a negative differential value will be
displayed if T1<T2.