8.0 RECOMMENDATIONS
Formaldehyde:
• Implement an exposure control plan as per OHS Reg 5.57 (2) – Designated Substances. This
regulation states that substances designated as ALARA substances must be substituted,
however, if it is not practicable then the employer must implement an exposure control plan
to maintain workers’ exposure as low as reasonably achievable below the applicable
exposure limit.
• It is recommended that the leaking spigot on the formalin container be discussed with the
company representative. The leaking spigot must be maintained in good operating condition
by the department in order to prevent leaks.
Nitrous Oxide:
• Management should continue investigations to reduce peak exposures to below the 5X
excursion limit of 125 ppm.
• Install local exhaust ventilation. Few solutions are presented in the literature as to the type of
local exhaust system that would decrease the peak exposures, however, it is possible that a
slotted hood located at the head of the bed might be an effective solution.
and/or
• Conduct a feasibility study to look at the costs and benefits of increasing the number of room
air changes per hour.