4.3.1. Leak Level
This is the concentration level in PPM that will activate a leak alarm condition.
1. Press the ENTER key to adjust the value.
2. Use the UP/DOWN cursor keys to modify the setting.
3. Press ENTER to accept the new entry or ESC to revert to the previous setting.
NOTE: The leak level value must be less than the spill level.
4.3.2. Spill Level
This is the concentration level in PPM that will activate a spill alarm condition.
1. Press the ENTER key to adjust the value.
2. Use the UP/DOWN cursor keys to modify the setting.
3. Press ENTER to accept the new entry or ESC to revert to the previous setting.
NOTE: The spill level value must be less than the evacuation level and greater than the
leak level.
4.3.3. Evacuation Level
This is the concentration level in PPM that will activate an evacuate alarm condition.
1. Press the ENTER key to adjust the value.
2. Use the UP/DOWN cursor keys to modify the setting.
3. Press ENTER to accept the new entry or ESC to revert to the previous setting.
NOTE: The evacuation level value must be greater than the spill level.
4.3.4. Re-Setting the Peak PPM Value
Pressing this key resets the current peak PPM level stored in memory and displays it at the top of the
screen.
4.4. Alarms
4.4.1. Functional Overview
If the PPM level for any zone exceeds its designated spill, leak, or evacuate thresholds, an alarm condition
will be created. Once the MZ monitor completes a measurement cycle in the affected zone the alarm
condition will be indicated. At that time the red ALARM LED on the MZ monitor will glow. Additionally, an
external alarm device may activate and an audible alarm may sound if those features have been enabled.
The next time the RD polls the affected monitor its red ALARM LED will glow. Additionally, an external
alarm device may activate and an audible alarm may sound if those features have been enabled.