Specifications

Halogen Gas Monitor – Multi Zone
6 Instruction 3015-5074
Safety Precautions
AC Power Supply
The HGM-MZ uses a universal power supply that is capable of accepting inputs of 100 to 240 VAC, 50/60 Hz.
The monitor’s power consumption is 20 Watts. It is highly recommended that the monitor be connected
directly to the AC power source, preferably on its own circuit with UPS or surge protection.
A switch or circuit breaker must be included in the building installation.
The switch must be in close proximity to the monitor and within easy reach of the operator.
The switch must be clearly marked as the disconnecting device for the equipment.
Protective Grounding
Under no circumstances should the HGM-MZ be operated without connection to a protective ground. Doing
so poses a potential shock hazard and is also a violation of electrical safety standards applicable to this type
of equipment.
Explosive Atmosphere
Do not operate this equipment in the presence of flammable liquids, vapors or aerosols. Operation of any
electrical instrument in such an environment constitutes a safety hazard.
Proper Exhaust Venting
It is imperative that the exhaust port on this instrument be properly vented as described in this manual.
Failure to do so may constitute a safety hazard.
Accessing the Interior of the Monitor
Extreme care should be exercised when accessing the interior of the monitor. Only qualified electrical
maintenance personnel should perform connections and adjustments. Always remove AC power before
opening the monitor’s enclosure.
Misuse & Modifications to the Instrument
The protection provided by the monitor may be impaired if the monitor is used in a manner not specified by
Bacharach, Inc. Modifications to this monitor, not expressly approved, will void the warranty.
In Case of Malfunction
Do not continue to use this equipment if there are any symptoms of malfunction or failure. In the case of such
occurrence, de-energize the power supply and contact a qualified repair technician or the nearest Bacharach
Service Center. Use ONLY the provided knockouts for electrical and communication wiring. Drilling into the
box will void the warranty.
Fusing
F1, F2: 1.0 A, 250 VAC, Type “F”
Installation Category
Installation Category II, Pollution Degree II, as defined by UL.
Altitude Limit
6,562 ft (2,000 m)
Cleaning
To clean the outside of the case use a DRY CLOTH. To avoid shock hazard and/or equipment
damage, DO NOT use soap and water.