Halogen Gas Monitor – Multi Zone
Instruction 3015-5074 41
The Diagnostic Screen
Navigating to the Diagnostic Screen
On the System Setup screen, select the Diagnostic option (DIAG).
Diagnostic Screen
01 ZONE 1 LOC
DET 4.2251V
AVE 4.2273V
ZERO 4.2267V
NOISE 0.0009V
AVEAU 0.00014
UM/L 0.000B 0.000AMB
PPM 0.7B 0.8STP
R134A NOW ZERO DELTA
BENCH 27.78C 27.80C -0.02C
BOX 29.97C 29.96C 0.01C
PRES 13.63PSIA -0.85 DP
VAC 11.34PSIA -3.20 DP
AMB 14.54PSIA 2.29DV
FAULT CODE=0000 0100 FEET
F
F
F
F
F
F
F
F
Indicates parameter is out of
tolerance & causing a fault condition.
Optical Sensor
Output Voltages
Receive port #bytes
Transmitport #bytes
Modbus function
Register address
Absorbance &
concentration data
Tempsat last purge
& difference from
current temps
Manifold - ambient
Vacuum - ambient
Manifold -vacuum
Programmed length
of tubing for active
zone
Diagnostic Screen Overview
The Diagnostic Screen contains sensor data and status information useful for trouble shooting various fault
conditions. An explanation of the information provided on each line of this screen is listed below, including
normal operating ranges.
DET: Detector Voltage – Peak-to-peak output of the IR sensor. In the absence of gas, this value
range is 3.900V to 4.500V.
AVE: Average Detector Voltage – Running average of the values displayed in line 1.
ZERO: Zero Voltage – IR sensor output that was stored during the last purge cycle and has the
same range as line 1.