Specifications
Ammonia Gas Monitor – Multi Zone
40
Instruction 3015-5113
Programming New Gasses
As new gasses come into use, the AGM-MZ allows the addition of these new gases to its gas library. At the
end of the gas library list is an option labeled CUSTOM for adding new gases. From the Calibration Screen
use the LEFT/RIGHT arrows to select CUSTOM from the list. Next, press ENTER, and use the UP/DOWN
arrow keys to enter the calibration factor. The selection of the matching gas and CAL factor is performed by
Bacharach by analyzing the new gas. Once the matching gas is determined, field calibration is possible by
using the same procedure as for other gasses. When the new gas entry is complete, setup the appropriate
zone for CUSTOM.
Program the instrument for a new gas as follows:
1. From the Calibration Screen, use the PREV GAS or NEXT GAS options to scroll through the list of
available choices until the CUSTOM option is displayed.
Custom Gas Screen
2. Select the CUSTOM option. Press ENTER to accept the new entry or ESC to revert to the previous
setting.
3. Enter the new CAL Factor as received from Bacharach.