Specifications

Ammonia Gas Monitor – Multi Zone
28
Instruction 3015-5113
Navigating to the 3
rd
Setup Screen
From AGM System Setup Screen #2, select the MORE option to go to AGM System Setup Screen #3.
Select the BACK option to return to AGM System Setup Screen #2.
AGM System Setup Screen #3
NOTE: The RS485 BAUD rate text will be replaced by “Service Timeout” when in Service Mode.
Service Timeout
Displayed in Service Mode. Sets the length of time the Service Mode is in effect. Stops the monitor for up to
300 minutes (5 hrs) to allow for servicing the unit.
1. Press the ENTER key to adjust the setting.
2. Use the LEFT/RIGHT cursor keys to move across the entry field and the UP/DOWN cursor keys to
modify the individual numbers.
3. Press ENTER to accept the new entry or ESC to revert to the previous setting.
Node Address
Each AGM on the network must has a distinct node address. The node address may be set from 1 to 64.
1. Press the ENTER key to adjust the setting.
2. Use the LEFT/RIGHT cursor keys to move across the entry field and the UP/DOWN cursor keys to
modify the individual numbers.
3. Press ENTER to accept the new entry or ESC to revert to the previous setting.
Password
This field is used to define a system password. The default setting is 000, which provides no password
protection.
1. Press the ENTER key to adjust the password.
2. Use the LEFT/RIGHT cursor keys to move across the entry field and the UP/DOWN cursor keys to
modify the individual alphanumeric characters.
3. Press ENTER to accept the new password or ESC to revert to the previous setting.