B.3.13. Sensor Data Register 0x0016h (22 Dec) (R, 82 Bytes)
Variable Type Length Description
Float
4
Manifold Pressure is PSIA
Float
4
Pressure sensor output Voltage
Float
4
Pressure with all value closed and pump on in PSIA
Float
4
Absolute Ambient Pressure in PSIA
Float
4
Enclosure Temperature in Degrees C
Float
4
Box temp sensor output voltage
Float
4
Optical bench temperature in Degrees C
Float
4
Bench temp sensor output voltage
Float
4
Optical bench temp in degrees C at last zero interval
Float
4
Box temp in degrees C at last zeroing
UI
2
Current peak to peak A/D counts from detector
Float
4
Current peak to peak voltage from detector
Float
4
Running average voltage from detector
Float
4
Voltage measured at last zeroing
Noise Float 4 Largest Change in running average
Float
4
Current absorbance value
Ave_AU Float 4 Running Average of absorbance level
Bench_PPM Float 4
PPM in bench based on zone gas selection (uncorrected for
pressure and temperature)
Float
4
PPM corrected to STP (1 atm, 25 deg C)
Float
4
Micromoles/liter in bench (uncorrected)
Ambient_UML Float 4 Micromoles/liter corrected to ambient pressure
B.3.14. Release Zone Hold Register 0x0017h (23 Dec) (W, 10 Bytes)
Variable Type Length Description
Rel_Hold
*
*
See description of STATUS REGISTER
B.3.15. Hold Zone Register 0x0018h (23 Dec) (W, 10 Bytes)
Variable Type Length Description
Zone_Hold
*
*
See description of STATUS REGISTER
B.3.16. MZ Hold Mode
The MZ monitor can be made to hold or “dwell” on a particular zone if necessary. The length of the hold
time is defined by the HOLD TIME parameter in the System Data Register.
Placing the MZ Monitor into hold mode:
1. Read the MZ Status Register (0x0011h)
2. Modify the content of the status register structure to change the MODE parameter to zone hold
mode. Active zone parameter to the zone you wish to hold
3. Send this updated status register structure back to the MZ using PRESET MULTIPLE REGISTER
COMMAND to the HOLD ZONE REGISTER (0x0018h).
Releasing the Zone Hold
1. Read the MZ monitor Status Register (0x0011h)