B.3.9.  Fault Code Table 
BIT  Fau lt Type  Code  Description 
15 
CLIPPING FAULT 
0x8000 
0x8000 Sensor output voltage too high 
14 
ZERO RANGE FAULT 
0x4000 
Sensor output voltage outside limits during purge 
13 
OVER R ANGE F AULT 
0x2000 
Gas concentration above measurable range 
12 
PURGE FLOW FAULT 
0x1000 
No flow detected during purge cycle 
11 
ZONE FLOW FAULT 
0x0800 
No flow detected during sample cycle 
10 
A2D F AULT 
0x0400 
Analog to digital converter not working   
9 
GAIN SET FAULT 
0x0200 
Digipot gain setting out of normal range  
8 
ZERO FILTER FAULT 
0x0100 
Charcoal filter (if used) needs replacement 
7 
CONFIG FAULT 
0x0080 
No zones are enabled  
6 
UNUSED FAULT 2 
0x0040 
5 
UNUSED FAULT 1 
0x0020 
4  LOOP FAULT  0x0010  Current loop is open   
3 
RS485 FAULT 
0x0008 
Communications  error  
2 
MANIFOLD P FAULT 
0x0004 
Pressure sensor readings are out of range 
1 
BENCH T FAULT 
0x0002 
Sensor temperature is out of range 
0 
BOX T FAULT 
0x0001 
Chassis temperature is out of range 
B.3.10. Zone Data Register 0x12xx (R/W, 78 Bytes) 
Each zone for an MZ has a separate zone data structure that is 78 bytes long. The zone number is the low 
order byte in the register address (i.e., Zone 1 data register = 0x1201h). 
Variable  Type  Length  Description 
Location 
C 
13 bytes 
13 byte array, Alpha Numeric Description or Name of Zone 
Flow OK  UC  1 byte 
Status of Flow check. Value of 1 indicates flow check is good. DO NOT 
MODIFY. 
Refrigerant 
Type  
UC  1 byte  See note 1 Below (DEFAULT = R134a) 
Distance 
UI 
2 bytes 
Zone Tubing Length (in feet) (DEFAULT = 100 feet [approx. 30.5 meters]) 
Zone Temp 
I 
2 bytes 
Average temperature at zone (degrees C) (DEFAULT = 25°C) 
Concentration 
Float 
4 bytes 
Last Measured concentrations (uM/L) DO NOT MODIFY 
Concentration2 
Float 
4 bytes 
Last Measured concentration (PPM) DO NOT MODIFY 
Alarm  Ack   UC  1 bytes 
Set value to 1 to acknowledge Alarm.  
NOTE: MZ will reset this byte to 0 when the Alarm byte 
(below) is = 0 and zone in alarm is sampled. If the alarm 
condition/byte increases (leak>>s pill or spill>>evac) the MZ 
will also reset this byte to 0. 
Alarm 
UC 
1 bytes 
Alarm Status; 0 = no alarm, 1 = leak; 2 = spill;  3 = evac.  
Leak Level 
UI 
2 bytes 
Level to trigger a leak alarm  (in PPM) (DEFAULT = 100) 
Spill Level 
UI 
2 bytes 
Level to trigger a spill alarm (in PPM) (DEFAULT = 300) 
Evac Level 
UI 
2 bytes 
Level to trigger a evacuate alarm (in PPM) (DEFAULT = 500) 
Peak PPM 
UI 
2 bytes 
Highest Recorded PPM in zone  
Peak Time 
TIM 
13 bytes 
Date an time of highest peak (see note 2 for format) 
Alarm Time 
TIM 
13 bytes 
Date and time of last alarm (see note 2 for format) 
Alarm Ack/Time 
13 bytes 
Log Interval 
UI 
2 bytes 
Number of minutes between Log entries (DEFAULT = 1440) 