6. Connect the tubing of the Calibration Kit to the GAS inlet of the an-
alyzer. Note that the gas fitting (Item 6) is only required on Fyrite
Pro Models 105 and 125.
7. Adjust the regulator for approximately 2 SCFH of excess flow (see
Figure 5.1).
8. After the analyzer has stabilized (2 to 3 minutes), use the and
buttons to set the displayed value to match the CO concentration
stamped on the gas cylinder.
9. Press ENTER to store the new calibration values and move to the
next calibration screen, or hold down ENTER for 2 seconds to store
the new calibration values and enter the Run Mode, or press the
I/O button to exit the Calibration Mode and turn OFF the analyzer
without saving the changes.
10. Disconnect tubing from analyzer and turn off gas flow.
Figure 5-1. Calibration Kit Hookup