Specifications

Instruction 24-9395
33
Fyrite Pro
4.13 Screen Descriptions
4.13.1 Warm-Up Screens
As soon as the analyzer is turned ON, a series of Warm-Up Screens are dis-
played. These screens show the analyzers model number (Pro 100/105/110/
120/125), the unit’s software revision, and finally the remaining warm-up
time counted down from either 10 or 60 seconds. If an O
2
sensor is installed,
then its current output level is also displayed. The 60 second countdown
only occurs when the analyzer contains an O
2
sensor, or if the analyzer con-
tains a CO sensor that has been set up for auto zero (refer to Section 3.5.3).
At the end of warm-up, the analyzer’s first display screen appears (refer to
Section 4.12).
Note: If any errors were detected during warm-up, the
Sensor Error Screen (Section 4.13.13) is displayed.
Tip: An O
2
error will occur when the oxygen sensor’s output
drops to between 80 and 90. Consider replacing the oxygen
sensor when its output level drops below 100.
4.13.2 CO Screen (Model 100)
The CO Screen is displayed after warm-up and shows the current CO level
in ppm. If the probe is being used, insert the probe tip into the area to be
sampled.
Front Panel Button Functions:
No effect
No effect
HOLD Freezes display (press ENTER to unfreeze)
ENTER Restarts testing after the HOLD button was pressed
LIGHT Toggles backlight ON/OFF
I/O Turns analyzer OFF (5 second delay)