Specifications

Page 1-4 Bacharach, Inc. Instructions 77-9045
INTRODUCTION CD3 AUDIT KIT
3. Replace fluid and filters in the calibration fluid
system.
NOTE: Save approximately 2 quarts of
calibration fluid from the same batch as
fluid in step 3. Store in a clean container to
use when checking viscosity (Section 2.2).
4. Make sure all calibration fluid lines are intact
and clean.
5. Replace any worn components such as belts,
pulleys, and couplings.
6. Check and fill, if necessary, the lube oil in the
air-driven hydraulic pump.
7. Make sure the water, air and electrical utilities
meet specifications listed in the Injector Cali-
brator CD3 Installation/Operation/Maintenance
Instruction (67-9319).
8. Operate the calibrator with an injector in place
for 30 minutes before beginning the audit. This
will circulate new fluid and remove any con-
taminants that may be in the lines.
9. Use the following checklist to complete the
preparation for the audit.