Specifications
Instructions 77-9045 Bacharach, Inc. Page 2-17
CD3 AUDIT KIT CUMMINS INJECTORS
6. While holding the HOLD TO OPERATE button,
push the BODY TO SEAT CLAMP button on
the right hand side operator’s control panel.
NOTE: The BODY TO SEAT clamp
monitor located on gauge panel will appear
green when the body to seat clamp is
engaged.
7. Release both buttons.
8. To unclamp the load cell, press the yellow
button on the left hand side of the operator’s
control panel.
Test
9. Turn on the clamp force and temperature
indicator.
10. Clamp the body to the seat.
11. Record the reading on the clamp force and
temperature indicator.
12. Unclamp the body from the seat.
13. Repeat steps 10, 11 and 12 three times. Each
reading should be a minimum of 167 pounds
force.
14. If readings are too low:
a. Check the gauge reading on the air regula-
tor located in back of calibrator. It should be
set for 60 to 65 psi.
b. Make sure there are no air leaks in the
lines supplying air to clamping cylinders or
solenoids.
c. Examine all affected parts and replace
damaged or worn components.
15. Unclamp the load cell assembly.
16. Turn off the clamp force and temperature
indicator.