Specifications

Instructions 77-9045 Bacharach, Inc. Page 2-5
CD3 AUDIT KIT CUMMINS INJECTORS
2. Immerse the Viscor cup and temperature
transducer in the fluid in the container.
3. Allow a minimum of one hour for the tempera-
ture of the fluid, cup, and transducer to stabi-
lize at room temperature.
4. Connect the temperature sensor cable to the
clamp force and temperature indicator.
Test
5. Follow the instructions supplied with the Viscor
cup.
6. Return contents of the container to the fuel
reservoir, clean the equipment and return the
items to the Audit Kit case.
7. Leave the temperature sensor connected to the
clamp force and temperature indicator. Turn off
the indicator.
2.3 CALIBRATION FLUID
TEMPERATURE TEST
Maintaining the correct calibration fluid tempera-
ture at the injector inlet is important because the
fluid temperature has a marked influence on
delivery.
Equipment Required
• Clamp force and temperature indicator
• Temperature transducer assembly
• Cummins, Flow Code 178A injector
Set-up
1. Turn the circuit breaker on the side of the
calibrator to ON.
2. Allow time for calibration fluid to reach the set
temperature.
NOTE: Allow the HEATER ON indicator
(located on front panel just below the
CALIBRATION FLUID TEMPERATURE
gauge) to cycle ON and OFF several times
before proceeding.
3. Connect the temperature transducer in line
with the fuel arm supply line hose.
Figure 2-7. Temperature Transducer
Figure 2-8. Transducer Installation
ËTEMPERATURE TRANSDUCER
ÏINDICATOR
ÍCUMMINS
æ178A INJECTOR
ËTEMPERATURE
æTRANSDUCER