Specifications
Page 1-2 Bacharach, Inc. Instructions 67-9319
INTRODUCTION INJECTOR CALIBRATOR CD3
Figure 1-2. Console Features
Feature Operation
_____________________________________________________________________
START METER Button Switches on flowmeter. Lights while meter is running.
(Flowmeter)
MM
3
/ STROKE Four-digit LED display with resolution of 0.1 mm
3
/stroke (Red
Display (Flowmeter) flashing dot indicates update in LED display).
STROKES Selector Selects 50 or 100 strokes per reading.
Switch (Flowmeter)
POWER “ON” When lit, indicates that the On/Off Switch is closed, and the pump
Indicator (light) for circulating calibration fluid and providing clamping pressure is
operating.
HOURS Meter 0 to 9999 hours (counts total hours while On/Off Switch is on).
CLAMP MONITOR/ Factory set to indicate proper plunger-to-body clamp force. When
TIMING, PLUNGER pointer points to the green segment of the dial, clamping force is
TO BODY gauge 380 lb ±10 lb.
Plunger to Body Clamp Lights for about 3 seconds during clamping operation.
Timing Indicator (light)
START MAIN Applies power to main drive motor. Starts camshaft rotation
DRIVE Button and the injector supply pump.
HEATER “ON” When lit, indicates power is being applied to calibration fluid
Indicator reservoir heater.
TEMPERATURE Permits manual setting of calibration fluid temperature.
Adjustment
(Potentiometer)
R A I L P R E S S U R E
G a u g e
P R E S S U R E
R E G U L A T O R
C A L I B R A T I O N F L U I D
T E M P E R A T U R E G a u g e
S T A R T M E T E R
P O W E R
I n d i c a t o r
S T R O K E S S e l e c t o r
S w i t c h
H O U R S M e t e r
C
L A M P M O N I T O R / T I M I N G
P
L U N G E R T O B O D Y G a u g e
P l u n g e r t o b o d y c l a m p
t i m i n g i n d i c a t o r
S T A R T M A I N
D R I V E
H E A T E R
I n d i c a t o r
T E M P E R A T U R E
A d j u s t m e n t
C L A M P M O N I T O R
B O D Y T O S E A T
M M / S T R O K E
D i s p l a y
3