MAINTENANCE AND TROUBLESHOOTING INJECTOR CALIBRATOR CD3
At the same time the fluid is changed, change the
suction and final stage filters.
1. Remove the back panels of the calibrator.
2. Drain the old fluid using the drain valves, one on
each side of the reservoir.
3. Lift the lid of the reservoir and inspect its bottom.
Wipe clean if necessary.
4. Close both drains of the reservoir.
5. Through the dipstick port, fill the reservoir to the
“F” mark on the dipstick with new fluid. Use
Bacharach part number 67-5598 (SAE J967).
NOTE: It’s a good idea to fasten a tag to the
reservoir giving the date the fluid was
changed.
6. Unscrew the suction filter cartridge from its
bracket and throw it away.
7. Replace the filter cartridge with a new one.
NOTE: Before installation fill the filter
cartridge with calibration fluid.
8. Using a wrench, remove the bolt from the top of
the final stage filter.
9. Remove the old filter and install a new one.
NOTE: Before installation fill the can with
new calibration fluid.
10. Reinstall the filter can.
11. Bleed excess air from the filter after starting the
calibrator.
NOTE: Fasten tags on both filters giving the
date they were changed.
12. Remove and clean flowmeter strainer.
Figure 4-4. Suction Filter
Figure 4-5. Final Stage Filter
Figure 4-6. Flowmeter Strainer
6 Suction filte
r
8 Final stage filter
12
Flowmeter straine
r