Specifications

Instructions 67-9319 Bacharach, Inc. Page 2-3
INJECTOR CALIBRATOR CD3 INSTALLATION
Figure 2-6. Air Connection
Figure 2-7. Adjust Air Pressure
Figure 2-8. Calibration Drains
Air
8. Connect an air line from the shop supply (95 to
125 psi) to the fitting at the rear of the calibra-
tor marked AIR (1/4 inch NPT).
9. Check regulator settings. The Primary Air
Pressure Regulator should be set at 90 ± 5 psi,
the Secondary Air Pressure Regulator must be
set at 60 +5 -0 psi (4.1 bar). The regulators are
behind the lower rear panel.
CAUTION:
HIGH PRESSURE CAN DAMAGE AIR
REGULATORS. Shop air supply must not
exceed 125 psi.
10. For regulator adjustments see Section 4.3.2
2.4 Filling Reservoirs
Calibration Fluid
1. Check that both drains of the calibration fluid
reservoir are closed.
2. Through the dipstick port, fill reservoir to the
“F” mark on the dipstick with new fluid. Use
Bacharach part number 67-5598 (SAE J967).
3. After a few minutes, check for leaks.
4. Place a tag on the reservoir giving date it was
filled.
Cambox
The cambox is filled with lube oil at the factory.
However, if filling is necessary:
5. Make sure cambox drain plug is in place. Drain
plug is accessible after left-hand cab inlet door
is open.
6. Fill the cambox with SAE 30W motor oil to
bring the level up to within 3-3/4 inch from the
top (or about 1-1/2 inch from the bottom).
Intensifier
7. The intensifier reservoir is filled at the factory
before shipping. If additional fluid is required,
see Section 2.6.
Lift Red Ring
9 Check Regulator Setting
8 Air in
9 Adjust Primary Air Pressure
to 90 psi
9 Adjust Secondary Air Pressure to 60 psi
1 Close Both Drains