® HI FLOW Sampler Natural Gas Leak Rate Measurement Instruction 0055-9017 Operation & Maintenance Rev.
WARRANTY Bacharach, Inc. warrants to Buyer that at the time of delivery this Product will be free from defects in material and manufacture and will conform substantially to Bacharach Inc.’s applicable specifications. Bacharach’s liability and Buyer’s remedy under this warranty are limited to the repair or replacement, at Bacharach’s option, of this Product or parts thereof returned to Seller at the factory of manufacture and shown to Bacharach Inc.
Hi Flow Sampler Contents Table of Contents 1 Introduction........................................................................................ 1 1.1 1.2 1.3 1.4 1.5 General Description................................................................................................. 1 Material Supplied .................................................................................................... 2 Why Use the Hi Flow Sampler?.........................................................................
Contents 4.1 4.2 4.3 4.4 4.5 4.6 Hi Flow Sampler View Calibration Dates ..........................................................................................43 Gas Calibration Equipment Setup .........................................................................43 Check Calibration Procedure .................................................................................44 Gas Calibration Procedure .....................................................................................
Hi Flow Sampler Introduction 1 Introduction 1.1 General Description The Hi Flow Sampler is portable, intrinsically safe, battery-powered instrument designed to determine the rate of gas leakage around various pipefittings, valve packings, and compressor seals found in natural gas transmission, storage, and compressor facilities. A component’s leak rate is measured by sampling at a high flow rate so as to capture all the gas leaking from the component along with a certain amount of surrounding air.
Introduction Hi Flow Sampler the range of 0.05 to 100% gas by volume. A second identical combustibles sensor channel measures the background CH4 level within the vicinity of the leaking component. The final element in the sampling system is a blower that exhausts the gas sample back into the atmosphere away from the sampling area.
Hi Flow Sampler 1.4 Introduction Control Unit Description The control unit consists of an 8 line by 20 character LCD and the four pushbuttons. Figure 1-1. Control Unit 1.
Introduction Hi Flow Sampler Notes: 4 Instruction 0055-9017
Hi Flow Sampler Technical Data 2 Technical Data Information Displayed .......... - Date and Time Battery voltage Sampling flow rate in cfm Background gas concentration in ppm or % by volume Leak concentration in ppm or % by volume Leak rate in cfm Percent difference between leak rate measurements #1 and #2 Latitude, Longitude, Altitude, and Time of Fix, when the optional GPS receiver is attached Display................................. 8 line by 20 character LCD Pushbutton Controls ............
Technical Data Hi Flow Sampler Natural Gas Sensor: Detection Method...............Catalytic oxidation / Thermal conductivity Range: Catalytic oxidation.........0 to 5% by volume methane Thermal conductivity .....5 to 100% by volume methane Accuracy............................±5% of reading or 0.02 % methane, whichever is greater Battery: Type ..................................Intrinsically Safe NiMH rechargeable pack Voltage ..............................4.8 V, max. Recharge Time ...................
Hi Flow Sampler Operation 3 Operation 3.1 Operating Precautions (Please Read) To ensure that the gas sensors are properly zeroed at startup, the Hi Flow Sampler must be turned ON in clean air (free of combustible gases or vapors). Calibrate the instrument about every 30 days to assure its accuracy. The frequency of calibration, however, may change depending on how often the instrument is used and the amount of gas that was sampled.
Operation 3.3 Hi Flow Sampler Sampling Hose & Attachment Connections Connect the sampling hose to the instrument’s leak-gas inlet by clamping the hose to the inlet as shown in Figure 3-2. Also connect the background gas sampling hose to the unit’s BACKGROUND inlet. Figure 3-2. Hose and Attachment Connections The standard sampling hose is 6 foot long.
Hi Flow Sampler 3.4 Operation Grounding WARNING: To prevent the possibility of a static discharge, the Hi Flow Sampler should be grounded while conducting a leak test. Attach the unit’s grounding clamp to the nearest earth ground. Tip: When not in use, the grounding clamp can be stored by attaching it to one of the shoulder straps. Figure 3-3.
Operation 3.5 Hi Flow Sampler Attachments A variety of attachments are available for the Hi Flow Sampler that simplifies the task of capturing the entire gas leak. Choose the appropriate attachment for the type of component being tested from the ones listed below, and then connect it to the end of the instrument’s main sampling hose. 3.5.
Hi Flow Sampler 3.5.3 Operation Capture Bag This reusable capture bag measures 36" x 36" and can be used to completely surround a component that may have multiple leak sources, or contains a leak source that is difficult to pinpoint or reach. Attach the Hi Flow Sampler’s main sampling hose to the bag by simply pushing the end of the hose into the bag’s hose coupling. A click should be heard when the hose has been successfully connected. Release the hose by pushing in the coupling’s release mechanism.
Operation Hi Flow Sampler This tool is useful in capturing leaks in valve stems and small fittings. 3.5.6 Claw Tool Use this tool when the leak source is spread out over a narrow area, such as a crack in a pipe, square flanges, and cylinder heads. 3.5.7 3-Port Background Gas Sampler Attaches to the end of the background gas sample hose to give better average background gas readings. 3.
Hi Flow Sampler Operation With the Hi Flow Sampler located in an area containing clean air (no combustible gases or vapors), turn ON the instrument by first pulling up the ON/OFF switch and then moving it to its ON position. WARNING: Turning ON the instrument in air contaminated with combustible gas will cause false readings to occur. Use one of Bacharach’s portable gas detection instruments to sample the surrounding area for levels of combustible gas before turning ON the instrument.
Operation Hi Flow Sampler Select the Basic or Expanded Menu Mode as follows: 1. From either the Basic or Expanded Main Screen, select Menu > Menu Mode. 1 1 C a S e M e / 2 2 / l i b r t F l n u M 0 4 0 9 : 4 5 : 3 0 a t i o n o w U n i t s o d e [ 0 0 0 1 ] ( M ) S - - 2. Highlight the desired menu mode, and then press the I/O key to select that mode. 1 1 / 2 2 / 0 4 0 9 : 4 5 : 3 0 B a s i c M e n u E x p a n d e d M e n u [ 0 0 0 1 ] ( M ) S - - 3. Press ESC to return to the Main Screen.
Hi Flow Sampler 3.8.1 Operation Basic Main Screen Date and Time: Date displayed as mm/dd/yy, while time is displayed in a 12hour format. The date and time are set per Section 5.8. Btry(V): Battery voltage Flow(lpm): Sample flow rate in either liters/min or cu.ft./min.
Operation 3.8.2 Hi Flow Sampler Expanded Main Screen Date and Time: Date displayed as mm/dd/yy, while time is displayed in a 12hour format. The date and time are set per Section 5.8. Save: Select to save all current measurement parameters in memory** Lat(°): Latitude in degrees* Menu: Select to display the first of several submenus (refer to Section 3.
Hi Flow Sampler 3.9 Operation Menu System Selecting Menu from either the Basic or Expanded Main Screen provides the user access to a system of submenus as shown in Figures 3-4 and 35. The items that are listed in the submenus depend on whether the instrument is in its Basic or Expanded menu mode. The functions listed are described in their associated sections of this manual. Figure 3-4. Basic Menu System Figure 3-5.
Operation Hi Flow Sampler 3.10 Battery Status Battery status is displayed as a single battery icon shown in the lowerleft hand portion of the LCD. Since the voltage curve of the battery pack drops off rapidly towards the end of its charge, the Battery status icon is useful only in showing when the battery pack is nearing depletion. It does not show the pack’s remaining percentage-of-charge. 4.0 volts and above (Shutdown imminent, less than 5 minutes of operating time remain) 3.
Hi Flow Sampler Operation 3.13 Menu Mode – Basic or Expanded Selecting Menu Mode from the Menu System allows the user to select whether only the most basic amount of information is displayed on the Main Screen (Basic Menu mode), or if all available information is displayed (Expanded Menu mode). Refer to Section 3.8 Selecting the Basic or Expanded Menu Mode for detailed information on how to select the desired menu mode, and learn about what type of information is displayed in each mode. 3.
Operation Hi Flow Sampler 3.14.2 Manual Measurement Mode 3.14.2.1 Manual 2-Stage Mode In the Manual 2-Stage mode the user starts a measurement at a high flow rate, waits for a stable leak rate reading, and then manually saves the readings using the Save function. The unit then automatically lowers the flow rate, at which time the user makes a second leak rate measurement.
Hi Flow Sampler Operation 3.14.2.2 Manual 1-Stage Mode In the Manual 1-Stage mode the user starts a measurement at a high flow rate and then waits for a stable leak rate reading to occur. During this time the flow rate can be changed using the Speed function on the Main Screen. After a stable leak rate reading is obtained, the user can manually save all final measured and calculated values in memory using the Save function on the Expanded Main Screen.
Operation Hi Flow Sampler 3.15 Access Records Note: Send Records can only be selected if operating in the Expanded Menu Mode. Refer to Section 3.8. Selecting Access Records from the Expanded Menu System allows a user to either: 1) send all saved records to a personal computer, 2) view the records on the unit’s LCD screen, or 3) erase all saved records. 3.15.
Hi Flow Sampler Operation d. Set Port Settings to: - Bits per second: 115200 - Data bits: 8 - Parity: None - Stop bits: 1 - Flow control: None Then click OK to display HyperTerminal’s main window. e. Click File > Save. This creates a file with an .ht extension using the name that was entered in Step b. 3. After HyperTerminal is running: click Transfer > Capture Text > Start to accept the default drive\directory\filename to which all received data will be stored as an ASCII text file.
Operation Hi Flow Sampler 3.15.2 View Records View the records stored in memory on the unit’s LCD screen as follows: 1. Access the Review Record Screen from the Expanded Main Screen by selecting Menu > Access Records > View Records. This screen allows the user to view a short summary of each record stored in memory. If no records are in memory, then the message “No Records Saved” briefly appears in the message line.
Hi Flow Sampler Operation 3.16 Importing Saved Data into a Spreadsheet Test records that were downloaded to a personal computer per Section 3.15.1 can be imported into spreadsheet programs that are capable of importing comma-delimited files. Note that each data record consists of 24 fields, some of which may be blank because of options not being installed. TABLE 3-1. COMMA-DELIMITED FIELDS 1 2 3 4 5 6 7 8 9 Label in Column Heading Record# Inst.
Operation Hi Flow Sampler TABLE 3-2. TYPICAL SPREADSHEET* Record# 1 2 Instr.Serial# 123456 123457 Back#1(%) 0.0350 0.0155 Flow#2(cfm) 7.4 7.3 Date#1(MM/DD/YY) 01/20/03 01/20/03 Leak#1(%) 15.1500 21.3250 Leak#1(cfm) 1.49 2.09 Back#2(%) 0.0375 0.0245 Leak#2(%) 21.2300 29.1800 Time#1(HH:MM:SS) 10:15:05 01:25:30 Date#2(MM/DD/YY) 01/20/03 01/20/03 Leak#2(cfm) 1.5 2.1 Barcode Symbology 031323078 56098 031323078 56099 GPS Latitude(deg) 40.5370 40.
Hi Flow Sampler Operation 3.17 Access Test IDs Note: Access Test IDs can only be selected if operating in the Expanded Menu Mode. Refer to Section 3.8. Saved test records can be identified (e.g., location of test, type of test, name of operator, etc.) by manually entering up to two lines of text, with each line containing a maximum of 20 alphanumerical characters. This information will be associated with all succeeding tests, and will appear when the test record is either viewed or printed.
Operation Hi Flow Sampler 3. Select the displayed Test ID by pressing the I/Okey, and then return to the Main Screen by pressing ESC twice. Observe that [0005] is now displayed at the bottom of the screen. 1 1 L e S p S a S t M e / 2 a k e e v e a r n u [ 0 2 / 0 4 0 9 : 4 5 : 3 0 ( N o t A v a i l . ) d ( L O < > H I ) t 0 0 5 ] ( A ) S - - 3.17.2 Edit Test IDs Edit a previously entered Test ID or to create a new one as follows: 1. First select the Test ID to edit per Section 3.17.
Hi Flow Sampler Operation 5. Press the ESC key twice to return to the Main Screen. 3.17.3 Send Test IDs The following describes how to download the current list of Test IDs to a computer that is running Windows® HyperTerminal as the communications program. If a different operating system and/or communications program is being used, then please consult the appropriate instruction manuals for those products. 1.
Operation Hi Flow Sampler 4. From the High Flow Sampler’s Expanded Main Screen, start the download process by selecting Menu > Access Test IDs > Send Test IDs. 1 1 S e E d S e E r / 2 2 / 0 4 0 9 : 4 5 : 3 0 l e c t T e s t I D s i t T e s t I D s n d T e s t I D s a s e T e s t I D s [ 0 0 0 4 ] ( A ) S - - Observe that the Hi Flow’s LCD displays the message “Sending record x of y @ 115.2kbps,” where “x” is the current record being sent and “y” is the number of records in memory.
Hi Flow Sampler Operation 2. Select Yes to erase all Test ID information, or select No to return to the previous screen. 1 1 / 2 2 / 0 4 E r a s e 0 9 : 4 5 : 3 0 A l l T e s t Y e s N o [ 0 0 0 4 ] ( A ) I D s ? S - - 3. Press ESC twice to return to the Main Screen.
Operation Hi Flow Sampler 3.18 Making a Leak Rate Measurement Before making a leak rate measurement, the user must decide whether to use . . . the Basic or Expanded Menu Mode (Section 3.8) lpm or cfm for the flow rate (Section 3.12) the Automatic 2-Stage, Manual 2-Stage, or Manual 1-Stage method of making a measurement when operating in the Expanded Menu Mode (Section 3.14) a Test ID when operating in the Expanded Menu Mode (Section 3.17) 3.18.
Hi Flow Sampler Operation 1 1 F l B a L e L e S p / 2 o w c k a k a k e e [ 0 S a m p l 2 / ( l ( % ( % ( l d ( 0 0 e 0 4 0 9 : 4 p m ) 2 1 6 ) 0 ) 1 5 p m ) 4 2 L O < 1 ] ( M ) T i m e - > 7 : . 5 . 8 . 0 . 5 > 0 2 3 0 0 0 H I ) 1 : 0 0 6. In the example below, the current flow rate is 216.5 lpm, the background gas concentration is 0.8%, the leak concentration is 15% gas, and the calculated leak rate is 42.5 lpm.
Operation Hi Flow Sampler 1 1 L e L e S p S a S t / 2 a k a k e e v e a r 2 / ( % ( N d ( 0 9 : 4 5 : 3 0 0 . 0 0 o t A v a i l . ) L O < > H I ) ) 0 4 t [ 0 0 0 1 ] ( A ) S - - 5. The user is now asked to choose a new or different Test ID. Selecting Yes causes the Access Records menu to appear (refer to Section 3.15). After selecting or creating a new Test ID, start the measurement by pressing ESC until the Main Screen appears.
Hi Flow Sampler Operation 7. After both measurements are complete, the values in the Main Screen are put on hold (as indicated by “H--” appearing at the bottom of the screen), allowing the user to make note of these values before turning off the instrument or proceeding to the next measurement. In the examples below, the measured leak concentration was 15% gas during the first measurement and 18% gas in the second, while the calculated leak rates were 42.5 and 40.8 lpm, respectively. A difference of 4%.
Operation Hi Flow Sampler 3.18.3 Leak Rate Measurement – Manual 2-Stage Mode Note: The Manual 2-Stage Mode can only be selected if operating in the Expanded Menu Mode. Refer to Section 3.8. 1. If not already done, place the instrument into its Manual 2-Stage mode. Refer to Section 3.14.2.1 Manual 2-Stage Mode. 2. Choose an attachment that will ensure the complete capture of the gas leak.
Hi Flow Sampler Operation 7. Once a stable leak measurement is obtained, highlight Save and press the I/Okey. Observe that the measurement is saved in memory, and that the flow rate is automatically lowered. In the example below, the measured leak concentration for measurement #1 was 15% gas, and the calculated leak rate was 42.5 lpm. 1 1 B a L e L e S p S a / 2 c k a k a k e e v e [ 0 S a m p l 2 / ( % ( % ( l d ( 0 4 0 9 : 4 0 ) 1 5 p m ) 4 2 L O < ) 0 0 1 ] ( M ) e T i m e - > 7 : . 8 . 0 .
Operation Hi Flow Sampler 3.18.4 Leak Rate Measurement – Manual 1-Stage Mode Note: The Manual 1-Stage Mode can only be selected if operating in the Expanded Menu Mode. Refer to Section 3.8. 1. If not already done, place the instrument into its Manual 1-Stage mode. Refer to Section 3.14.2.2 Manual 1-Stage Mode. 2. Choose an attachment that will ensure the complete capture of the gas leak.
Hi Flow Sampler Operation 1 1 F l B a L e L e S p / 2 o w c k a k a k e e 2 / ( l ( % ( % ( l d ( 0 4 0 9 : 4 p m ) 0 ) 0 ) 0 p m ) 0 L O < [ 0 0 0 1 ] ( M ) 5 : . 0 . 0 . 0 . 0 > 3 0 0 0 Speed Bar H I ) - 1 - 7. Once a stable leak measurement is obtained, all current readings can be saved in memory by highlighting Save and pressing the I/Okey. In the example below, the current measured leak concentration is 15% gas, and the calculated leak rate is 42.5 lpm.
Operation Hi Flow Sampler 3.19 Message Line Various messages will appear at the bottom of the LCD that describe what functions are currently being performed or errors that have been detected. The following list describes the meaning of these messages. Note that detailed explanations of the error messages can be found in Section 6.2 Troubleshooting Guide. Calibration Passed Calibration procedure was successful. Calibration Failed Error – The calibration procedure was not successful. Refer to Section 6.2.
Hi Flow Sampler Operation Check Zeroing Bypassed Error – The ESC key was pressed during start up, thus bypassing the normal sensor-zeroing process. Refer to Section 6.2. Check Leak Rate Difference Error–The calculated leak rate difference between measurements #1 and #2 during an automatic or manual leak test was greater than 100%. Refer to Section 6.2. Check Background Greater Than Leak Error–The measured background gas level was greater than the measured leak gas concentration. Refer to Section 6.
Operation Hi Flow Sampler The GPS data is saved along with all other measured and calculated values at the conclusion of an automatic test, or when Save is selected during a manual test. 3.21 Barcode Scanner Data Note: Barcode scanning can only be performed if operating in the Expanded Menu Mode. Refer to Section 3.8. When a barcode scanner is connected to the instrument’s SCANNER port, the instrument acknowledges this connection by displaying at the bottom left corner of the LCD.
Hi Flow Sampler Calibration 4 Calibration It is recommended that the Hi Flow Sampler be calibrated by your nearest Bacharach Service Center every 30 days. Calibration, however, can be performed in the field if your facility has the necessary equipment and qualified personnel to perform the procedures described in the following sections. The following procedures describe how to: 4.
Calibration Hi Flow Sampler Figure 4-1. Calibration Equipment Setup 4.3 Check Calibration Procedure This procedure checks the calibration of both the leak-gas sensor and the background-gas sensor using 2.5% CH4. After setting up the calibration equipment as described in Section 4.2, proceed to check the instrument’s sensor calibration as follows: 1.
Hi Flow Sampler Calibration 5. Disconnect the hose from the BACKGROUND inlet port and connect it to the GAS inlet port. 6. Again wait several minutes for the gas reading to stabilize. At this time the gas concentration shown on the display should match the concentration stamped on the calibration gas cylinder. 1 1 / 2 2 / 0 4 1 0 : 4 5 : 3 0 B a c k ( % ) L e a k ( % ) [ 0 0 0 5 ] 0 . 0 0 2 . 5 0 ( M ) S - - 7.
Calibration Hi Flow Sampler 5. Apply the appropriate level of calibration gas from the calibration equipment to the instrument’s GAS or BACKGROUND inlet port. (In the example above, apply 2.5% CH4 to the BACKGROUND port.) Then press the I/Okey to start the calibration process. Observe that the gas-sampling-pump motors start and the calibration screen appears. 1 1 / 2 2 B a c k ( A p p l ( , = E N T E R = E S C = [ 0 0 / 0 % ) % ) A C E 0 5 4 1 0 : 4 0 2 d j . A p a l i b r a x i t ] ( M ) 5 : . 0 .
Hi Flow Sampler 4.5 Calibration Calibration Amplifier Gain Check Use this procedure to display the gain values of the calibration amplifier circuits in hexadecimal. This procedure is useful to service technicians for troubleshooting purposes. 1. From the Main Screen select Menu > Calibration > Calibration Gains to display the Calibration Gains Screen.
Calibration Hi Flow Sampler Notes: 48 Instruction 0055-9017
Hi Flow Sampler Maintenance 5 Maintenance Routine maintenance of the Hi Flow Sampler consists of the following: 5.1 Charging the battery pack Replacing the internal filters Replacing the gas sensors Verifying flow rate Setting the time and date Resetting the microprocessor Electrostatic Discharge Precautions CAUTION: Electrostatic discharge (ESD) can destroy sensitive electronic components inside the Hi Flow Sampler when its cover is removed.
Maintenance 5.2 Hi Flow Sampler Charging the Battery Pack WARNING: Explosion hazard! DO NOT charge the battery pack in an unsafe atmosphere. 1. Unplug battery pack from the instrument’s power cord; then remove battery pack from backpack. 2. Interconnect the supplied power supply, power cord, and charger as shown in Figure 5-1. Then plug the output of the charger into the battery pack’s charger jack.
Hi Flow Sampler 5.3 Maintenance Removing the Cover When instructed to do maintenance inside the instrument, remove its protective cover as follows: 1. Do the following before removing the instrument from the backpack: - Disconnect battery pack - Unscrew metal-cage diffuser from the instrument’s gas exhaust port - Disconnect ground clamp wire 2. Remove instrument from backpack. 3. Remove screws from cover as shown in Figure 5-2; then pull cover away from unit. Figure 5-2.
Maintenance 5.4 Hi Flow Sampler Replacing the Internal Filters Three internal filters protect the instrument from being contaminated by dust and dirt. Replace these filters when they look contaminated. A good indication that a filter needs replaced is when the instrument’s gasresponse time increases. 5.4.1 Leak-Gas Filter The leak-gas filter is located between the sampling port on the flowmeasurement tube and its associated solenoid valve on the printed circuit board. See Figure 5-3.
Hi Flow Sampler 5.4.2 Maintenance Leak-Gas Sensor Filter The leak-gas sensor filter is located inside the flow block of the leak-gas sensor assembly. See Figure 5-4. Equipment Required: Medium Phillips Head Screwdriver Replacement Filter (P/N 0055-0045) Procedure: 1. Remove cover per Section 5.3. 2. Remove tubing from inlet of sensor assembly. 3. Unscrew the filter chamber’s gas inlet fitting; then pull out old filter and discard. 4.
Maintenance 5.5 Hi Flow Sampler Gas Sensor Replacement Replace a gas sensor when it fails to calibrate per Section 4 Calibration. The gas sensors are located inside their associated flow blocks of the leak-gas and background-gas sensor assemblies. See Figure 5-5. Figure 5-5. Leak-Gas & Background Gas Sensor Assemblies Equipment Required: Medium Phillips Head Screwdriver Replacement Sensor (P/N 0055-0040) Procedure: 1. Remove cover per Section 5.3. 2.
Hi Flow Sampler 5.6 Maintenance Pump Replacement The pump located in either the leak-gas or background-gas sensor assembly (see Figure 5-5) may become inoperative due to water or solid particles getting into the valves. This will require replacement of the pump to correct. Equipment Required: 5/64" Allen Wrench Replacement Pump Assembly (P/N 0055-0061) Procedure: 1. Remove cover per Section 5.3. 2. Back off the pump-bracket screw using a 5/64" Allen wrench; then rotate bracket arm to free pump.
Maintenance 5.7 Hi Flow Sampler Flow Rate Verification This procedure verifies the instrument’s ability to provide sufficient airflow to capture the entire gas leak, and to verify the accuracy of the instrument’s calculated flow rate by using the optional external flow meter (P/N 0055-0270). Important: To assure accurate flow readings, conduct the following procedure in clean air, free of any combustible gases. 1. Install the flow meter onto the Hi Flow Sampler’s gas intake as shown in Figure 5-7. 2.
Hi Flow Sampler Maintenance Figure 5-8. Flow Rate Validation Graph 5. With a fully charged battery pack, the measured flow rate should be greater than 9 cfm. If not, perform the following (see Figures 5-9 and 5-10): Clean the gas sample inlet screen. Ensure that the exhaust port is not obstructed.
Maintenance Hi Flow Sampler Figure 5-9.
Hi Flow Sampler Maintenance Figure 5-10. Plumbing Diagram 5.8 Setting Date and Time The following describes how to set the date and time as displayed on the Hi Flow Sampler’s LCD using a personal computer that is running Windows® HyperTerminal as the communications program. If a different operating system and/or communications program is being used, then please consult the appropriate instruction manuals for those products. 1.
Maintenance Hi Flow Sampler - If HyperTerminal has never been run in association with the Hi Flow Sampler, then a new HyperTerminal connection needs to be made as follows: a. Select Start > Run. In the Run box type “hypertrm.exe” and click OK. The New Connection dialog box should appear. b. Type in a New Connection Name such as “HiFlow”. Click OK to select the default icon (if desired, select a different icon before clicking OK). The Connect To dialog box should appear. c.
Hi Flow Sampler Troubleshooting 6 Troubleshooting 6.1 Instrument Repair It is recommended that field repair of the Hi Flow Sampler be limited to: Cleaning gas-inlet screen Replacing internal filters Replacing sensors Replacing sensor assembly pump Simple checks of printed circuit board connectors All other repairs should be performed by an authorized Bacharach Service Center.
Troubleshooting Hi Flow Sampler TABLE 6-1. TROUBLESHOOTING GUIDE (Cont.) Instrument won’t respond when a control unit key is pressed. Microprocessor needs reset. Disconnect battery pack for several seconds. Low flow rate reading with fully charged battery. Flow path obstructed. Check flow path and verify flow rate per Section 5.7. Slow response time. Internal filters are dirty. Replace filters per Section 5.2. Battery power does not last at least 4.5 hours. Battery pack defective. Replace.
Hi Flow Sampler Troubleshooting TABLE 6-1. TROUBLESHOOTING GUIDE (Cont.) “Check Background Offset” message appears at bottom of display. The background-gas sensor did not zero properly during start-up. Switch OFF instrument and restart in an area with clean air. Replace sensor if necessary. “Check Leak Offset” message appears at bottom of display. The leak-gas sensor did not zero properly during start-up. Switch OFF instrument and restart in an area with clean air. Replace sensor if necessary.
Troubleshooting Hi Flow Sampler Notes: 64 Instruction 0055-9017
Hi Flow Sampler Parts & Service 7 Parts & Service 7.1 Replacement Parts Complete Hi Flow Sampler Assembly ...................................... 0055-8020 Standard Attachments: 3-Port Background Gas Sampler ....................................... 0055-0271 Bellows Tool ...................................................................... 0055-0259 Beveled Nozzle Tool: 6.5" ............................................................................. 0055-0213 24" ........................................
Parts & Service 7.3 Hi Flow Sampler Service Centers United States Bacharach Sales/Service Center 621 Hunt Valley Circle New Kensington, PA 15068 Phone: 724-334-5051 Fax: 724-334-5723 Email: help@mybacharach.com Canada Bacharach of Canada, Inc. 20 Amber St. Unit #7 Markham, Ontario L3R SP4 Canada Phone: 905-470-8985 FAX: 905-470-8963 E-mail: bachcan@idirect.
Headquarters: 621 Hunt Valley Circle, New Kensington, PA 15068 Ph: 724-334-5000 • Fax: 724-334-5001 • Toll Free: 1-800-736-4666 Website: www.mybacharach.com • E-mail: help@mybacharach.com Printed in U.S.A.