Specifications

Single Zone Gas Monitor
34 P/N: 3015-4256 Rev 13
APPENDIX B. AGENCY APPROVALS
DECLARATION OF CONFORMITY
The manufacturer of the
products covered by this
declaration:
Bacharach, Inc.
621 Hunt Valley Circle
New Kensington, PA 15068
Year conformity is declared: 2010 (EN 14624), 2008 (all others)
Product(s): Refrigerant Monitors
Model(s): HGM-SZ and HGM-MZ (Single and Multi Zone Refrigerant Monitors) (EN 14624 only)
HGM-SZ, AGM-SZ, CO2-SZ, HGM-MZ, AGM-MZ, CO2-MZ and MZ-RD.
The undersigned hereby declares that the above referenced products are in conformity with the provisions of the following
standard(s) and is in accordance with the following directive(s).
Standard(s):
EN 14624 Performance of mobile leak detectors and of room controllers of halogenated refrigerants
EN 61326-1:2006 Electrical equipment for measurement, control, and laboratory useEMC requirements: Part 1 General Requirements
EN 61010-1:2001 Safety requirement for control, measurement, and laboratory equipment
Directive(s):
2004/108/EC EMC Directive
2006/95/EC Low Voltage Directive
Signature:
Name: Doug Keeports
Title: VP of Product Development
Date: 06 December 2010
The technical documentation file required by this directive is maintained at the corporate headquarters of Bacharach, Inc.
European Standard EN14624: 2005 testing with R-134a (Halogen Gas Monitor Only).
Min. Sensitivity Threshold 1 ppm
Max. Sensitivity Threshold (within 1 ppm ± 10% reading accuracy range) 1,000 ppm
Max. Sensitivity Threshold within Instrument Reading Range 10,000 ppm
Zeroing Time from 1,000 ppm* <12 seconds
Zeroing Time from 10,000 ppm* <25 seconds
Reaction Time for detection of minimum threshold* ≤12 seconds
Min. Sensitivity threshold once max. threshold has been measured 1 ppm
Repeatability at 500 ppm ± 1% of reading
* Based on a minimum tubing length of one foot