Specifications
Maintenance H-SEM
16 Instruction 3015-4584
4.2 Internal Particulate / Hydrophobic Filter Replacement
Items Required:
• Replacement filter (07-1654)
• Medium Phillips head screwdriver
Procedure:
1. Disassemble the monitor’s metal chassis per Section
4.1.
2. See illustration below and locate the internal particulate / hydrophobic filter.
3. Pull off tubing from both ends of filter and remove filter from instrument.
4. Attach tubing to new filter, being sure to position the filter as shown in the illustration.
5. Reassemble monitor.
4.3 Flashing Memory
An RS-232 connector, located on the top panel, is used to flash the monitor’s memory. This connector is for
factory and service technician use only.