Owner's Manual

12
** Solution # corresponds to Cause # **
Problem No response to Calibration
Reference bottle
Cause
1
Heat Adjustment set too low or
bottle is empty
2
No air flow (indicator ball in
probe doesn’t float)
3
Sensor exposed to excessive
amounts of halogen gas
4
Water in probe
Solution
1
Readjust heater (see “Heater
Adjustment” section) or replace
bottle (3015-0864)
2
Replace filter in probe tip
(3015-0784). Check air pump
operation.
3
Move probe to clean atmo-
sphere for several minutes while
sensor purges itself.
4
Turn unit off and disconnect
probe from chassis. Remove
screws and take out chassis.
Look at the underside of the leak
detector and follow the probe
to the air pump. Remove this
hose from the pump. Blow clean
air (5 psi) into the probe tip for
one or two minutes. Reassemble
unit and replace the filter
(3015-0784).
Problem Probe tip lamp does
not flash
Cause Burned out or damaged probe
lamp
Solution Return for repair