Owner's Manual

GDX-350 Sensor Transmitter
P/N: 5600-9001 Rev 2 47
Memory ASCII Strings
VARIABLE ALIAS READ FUNCTION CODE WRITE FUNCTION CODE
User Info Chan 1
1
40401-40408 3 NA
User Info Chan 2
1
40409-40416 3 NA
Chan 1 ASCII Reading
2
40417-40419 3 NA
Chan 2 ASCII Reading
2
40420-40422 3 NA
EUNITS Chan 1
3
40423-40427 3 NA
EUNITS Chan 2
3
40428-40432 3 NA
1 16 ASCII characters (2 per register) assigned to the unit identifier read as bytes.
2 6 ASCII characters (2 per register) reflecting the display readout.
3 10 ASCII characters (2 per register) assigned to the engineering units read as bytes.
Byte Variables
VARIABLE ALIAS READ FUNCTION CODE WRITE FUNCTION CODE
PreAmp/Gain Ch1
1
40433 3 NA
PreAmp/Gain Ch2
1
40434 3 NA
1 2 bytes representing Pre Amp (HiByte) and PGA (LoByte) settings.
Firmware Version
2 ALIAS READ FUNCTION CODE WRITE FUNCTION CODE
Version
1
40435-40436 3 NA
1 4 ASCII characters (2 per register) reflecting the firmware version.
Memory Reals
NOTE: Real value represents float value without the decimal point such as 123.4 is returned as 1234.
Decimal divisor is returned as 1, 10, 100, or 1000 for decimal position of 1, 2, 3, or 4, where
123.4 would return the value 10.
VARIABLE ALIAS READ FUNCTION CODE WRITE FUNCTION CODE
Chan 1 Cal Zero Real 41001 4 NA
Chan 1 Cal Zero Divisor 41002 4 NA
Chan 1 Cal Span Real 41003 4 NA
Chan 1 Cal Span Divisor 41004 4 NA
Chan 1 Zero Real 41005 4 NA
Chan 1 Zero Divisor 41006 4 NA
Chan 1 Span Real 41007 4 NA
Chan 1 Span Divisor 41008 4 NA
Chan 1 Fault Real 41009 4 NA
Chan 1 Fault Divisor 41010 4 NA
Chan 1 Alarm 1 Real 41011 4 NA
Chan 1 Alarm 1 Divisor 41012 4 NA