Specifications

GDC-350 Operation Manual
5909-9000 Rev 3 31
CALIBRATING THE NULL (ZERO) (remote catalytic or SS sensor)
Step Calibrating the Null (Zero) Procedure
1 Attach a regulator to a cylinder of zero air.
2
Attach the flow adapter and open the regulator valve fully to allow zero
air to flow over sensor. Use a slight twisting motion as you gently push
the calibration adapter into the sensor opening. If the calibration adapter
is hard to insert, moisten the “O” ring seal slightly then try re-inserting it.
3
Move the jumper on J6 to P3 and move jumper on J7 to P3. The audible
alarm beeps once for confirmation. The system is now starting the zero
calibration stabilization time (30 seconds).
4
The instrument will beep twice when the calibration begins. Leave the
zero air flowing over the sensor until the GDC-350 beeps three times
indicating the procedure is finished (1 minute).
5
Move the jumpers from J6 and J7 back to their P1 resting positions and
remove the zero air (or nitrogen).
CALIBRATING THE SPAN (remote catalytic or SS sensor)
Step Calibrating the Span Procedure
1 Attach a regulator to a cylinder of span gas.
2
Insert the calibration adapter into the sensor opening on the front of the
enclosure. Use a slight twisting motion as you gently push the calibration
adapter into the sensor opening. If the calibration adapter is hard to
insert, moisten the “O” ring seal slightly then try re-inserting it.
3 Open regulator valve fully and allow span gas to flow over sensor.
4
Move the jumper on J6 to P3 and move the jumper on J7 to P4. The
audible alarm beeps once for confirmation. The system is now starting
the span calibration stabilization time (30 seconds).
5
The instrument will beep twice when the calibration begins. Leave the
span gas flowing over the sensor until the GDC-350 beeps three times
indicating the procedure is finished (2 minutes).
6
Move the jumpers from J6 and J7 back to their P1 resting positions and
remove the span gas.