Specifications

Instruction 0024-9435
8
Fyrite Tech
3.4 Front Panel Pushbuttons
Note that a pushbutton may perform several functions, depending on what
screen is being displayed at the time.
I/O Turns analyzer ON/OFF. There is a 60 second warm-up and
a 5 second turn-off-delay period.
Places the analyzer into either its Setup or Calibration Mode
when used in conjunction with the ENTER or HOLD button.
Aborts the CO purge function during turn-off.
Scrolls up through the display screens during a test.
Causes the displayed value to increase or change while in
the Calibration or Setup Mode.
Scrolls down through the display screens during a test.
Causes the displayed value to decrease or change while in
the Calibration or Setup Mode.
ENTER Enters the Run Mode (starts a combustion test) from the
Fuel Select Screen.
Displays the Fuel Select Screen when held down for 2 sec-
onds while in the Run Mode.
Sets up the analyzer to be placed into its Calibration Mode
when held down with the analyzer OFF. (Used in conjunc-
tion with the I/O button.)
Stores the displayed value and automatically steps to the
next screen when pressed during calibration or setup.
Places the unit into its Run Mode when held down for 2 sec-
onds while in the Calibration Mode.
Aborts turn-off and keeps the analyzer turned ON when
pressed during the 5 second turn-off-delay period.
HOLD Places the analyzer on hold and freezes the values in all
Run Mode Screens during a test, allowing the operator to
scroll through the displays and view all test values at that
point in time. Pressing ENTER resumes testing.
Sets up the analyzer to be placed into its Setup Mode when
held down with the analyzer OFF. (Used in conjunction with
the I/O button.)
Sends combustion test data to a printer when pressed twice
with the pump running (Fyrite Tech 60 only).
LIGHT Toggles the back light ON and OFF.