Fyrite Tech
3.4 Front Panel Pushbuttons
Note that a pushbutton may perform several functions, depending on what
screen is being displayed at the time.
I/O • Turns analyzer ON/OFF. There is a 60 second warm-up and
a 5 second turn-off-delay period.
• Places the analyzer into either its Setup or Calibration Mode
when used in conjunction with the ENTER or HOLD button.
• Aborts the CO purge function during turn-off.
• Scrolls up through the display screens during a test.
• Causes the displayed value to increase or change while in
the Calibration or Setup Mode.
• Scrolls down through the display screens during a test.
• Causes the displayed value to decrease or change while in
the Calibration or Setup Mode.
ENTER • Enters the Run Mode (starts a combustion test) from the
Fuel Select Screen.
• Displays the Fuel Select Screen when held down for 2 sec-
onds while in the Run Mode.
• Sets up the analyzer to be placed into its Calibration Mode
when held down with the analyzer OFF. (Used in conjunc-
tion with the I/O button.)
• Stores the displayed value and automatically steps to the
next screen when pressed during calibration or setup.
• Places the unit into its Run Mode when held down for 2 sec-
onds while in the Calibration Mode.
• Aborts turn-off and keeps the analyzer turned ON when
pressed during the 5 second turn-off-delay period.
HOLD • Places the analyzer on hold and freezes the values in all
Run Mode Screens during a test, allowing the operator to
scroll through the displays and view all test values at that
point in time. Pressing ENTER resumes testing.
• Sets up the analyzer to be placed into its Setup Mode when
held down with the analyzer OFF. (Used in conjunction with
the I/O button.)
• Sends combustion test data to a printer when pressed twice
with the pump running (Fyrite Tech 60 only).
LIGHT Toggles the back light ON and OFF.