Specifications

Instruction 24-9388
42
Fyrite Pro
5.8 Water Trap / Filter Maintenance
The Water Trap / Filter Assembly removes water condensate from the gas
sample, and also prevents soot from contaminating the internal compo-
nents of the analyzer.
Drain the water condensate after every test.
Procedure:
1. Pull off the Inlet End Cap using a slight twisting motion.
2. Pour out all water condensate; then reassemble trap.
Replace the Filter Element when dirty.
Material Required: • Filter Element, P/N 07-1644
Procedure:
1. Pull off the Outlet End Cap using a slight twisting motion.
2. Remove and discard old filter.
3. Install new filter; then reassemble trap.
Figure 5-3. Water Trap / Filter Assembly
OUTLET
FILTER ELEMENT
To re place Filter Elemen t,
pull off Outlet End Cap using
a slight twisting motion.
To empty Water Trap,
pull off Inlet E nd Cap using
a slight twisting motion.
INLET