Specifications
Instruction 24-9388
40
Fyrite Pro
8. Press ENTER to store the new calibration values and move to the
next calibration screen, or hold down ENTER for 2 seconds to store
the new calibration values and enter the Run Mode, or press the
I/O button to exit the Calibration Mode and turn OFF the analyzer
without saving the changes.
9. Disconnect tubing from analyzer and turn off gas flow.
Figure 5-1. Calibration Kit Hookup
1
2 3 4 5
LEAVE TOP
PORT OPEN
Parts Shown:
1. Gas Cylin der
2. Regulator*
3. Tubing*
4. Tee*
5. Flowmeter*
6. Fitting, Gas*
* Contained in Calibration Kit
3
6