5.6 CO Sensor Zero & Calibration
(Model 125E only)
Material Required: • Cylinder of 100 ppm (P/N 51-1994) or
500 ppm (P/N 24-0492) CO calibration gas
• Calibration Kit, P/N 24-7059
To improve the accuracy of the CO reading, we suggest that if the analyzer
will be primarily used for flue gas testing, then calibrate using 500 ppm
CO. If the analyzer, however, will be primarily used for ambient testing,
then calibrate using 100 ppm CO.
Procedure:
1. With the analyzer sampling fresh air, enter the Calibration Mode as
described in Section 5.1. Then repeatedly press the ENTER button
until “CO” appears in the upper-left side of the display.
0
CAL
CO
p
p m
2. Allow the pump to run and sample fresh air for at least 1 minute.
3. Use the and buttons to set the displayed value to 0 ppm.
4. Do one of the following:
a. End this procedure and save the new zero value by holding down
the ENTER button for 2 seconds; after which, the analyzer enters
its Run Mode.
b. Continue with Step 5 to span the CO sensor on a known concen-
tration of carbon monoxide.
5. Set up the Calibration Kit with 100 or 500 ppm CO as described in
the instructions supplied with the kit.
6. Connect the tubing of the Calibration Kit to the GAS inlet of the
analyzer; then adjust the regulator for approximately 2 SCFH of
excess flow (see Figure 5.1).
7. After the analyzer has stabilized (2 to 3 minutes), use the and
buttons to set the displayed value to match the CO concentration
stamped on the gas cylinder.