Specifications
Instruction 24-9388
35
Fyrite Pro
5.0 CALIBRATION & MAINTENANCE
Important: Fresh batteries should be installed, or use the
optional AC adapter to power the analyzer during calibra-
tion. The unit should be allowed to stabilize at room temper-
ature for at least 2 hours before proceeding with calibration.
To maintain accuracy as listed in the Technical Character-
istics Section of this manual, the standards used must be
at least four times as accurate as the stated accuracy of the
Fyrite Pro.
5.1 Entering the Calibration Mode and
Testing the Display Segments
1. With the analyzer turned OFF, place the unit in fresh, ambient air;
then press and hold down the ENTER button.
2. Press the I/O button and release it. Observe that all LCD segments
are turned ON.
3. Release the ENTER button. Observe the unit’s model number and
software version are displayed. The word “CAL” is then displayed
while the unit warms up and counts down from 60 seconds.
At the end of 60 seconds, the first calibration screen is displayed.
Note: The order in which the calibration screens are
displayed depends on whether the CO channel is set up for
automatic or manual zero (Section 3.5.3).
Calibration Screen Display Order
CO Auto Zero: Ambient, TL, TA, O
2
, CO, Pressure
CO Manual Zero: CO, Pressure, Ambient, TL, TA, O
2