Specifications

Instruction 24-9388
31
Fyrite Pro
4.13.7 qA & Excess Air Screen
The qA reading is a calculation of stack loss in percent.
The EXCESS AIR (LAMBDA) reading is the calculation of the percentage
of extra air that is available in the combustion chamber above the theoreti-
cal amount needed for perfect combustion.
Front Panel Button Functions:
Displays next screen (refer to Section 4.12)
Displays previous screen (refer to Section 4.12)
HOLD Freezes display (press ENTER to unfreeze)
ENTER Hold for 2 seconds to display Fuel Screen
LIGHT Toggles backlight ON/OFF
I/O Turns analyzer OFF (5 second delay)
4.13.8 TL & Eta Screen
The TL reading is the combustion-air temperature that is used for ef-
ficiency and qA calculations. This temperature is measured by either the
optional T-AIR thermocouple on a continuous basis, or by the T-STACK
thermocouple during start-up (refer to Section 4.2.1). If neither thermo-
couple is installed when the analyzer is turned ON, then this temperature
reading comes from a temperature sensor located inside the analyzer.
The Eta reading is the calculated percent efficiency as referenced to a low
heating value.
Front Panel Button Functions:
Displays next screen (refer to Section 4.12)
Displays previous screen (refer to Section 4.12)
HOLD Freezes display (press ENTER to unfreeze)
ENTER Hold for 2 seconds to display Fuel Screen
LIGHT Toggles backlight ON/OFF
I/O Turns analyzer OFF (5 second delay)