Specifications

Instruction 24-9388
6
Fyrite Pro
3.3 Connecting the Probe and Thermocouple
Attach the probe and hose assembly to the analyzer as follows (Figure 3-2):
1. Push the yellow banded quick-connect Flue Gas Hose (giving a
slight twist) onto the GAS inlet fitting.
2. Push the blue banded quick-connect Draft Hose (giving a slight
twist) on the “+” pressure fitting.
3. Push the Flue Gas Thermocouple into the T-STACK jack
Important: DO NOT force the thermocouple connector into
its jack. The connection tabs are different sizes, allowing the
connector to fit in only one way.
4. Push the optional combustion-air thermocouple into the T-AIR jack.
Note: The optional combustion-air thermocouple provides
a convenient way to measure the burner’s combustion-air
temperature when its temperature is not the same as room
air. If this thermocouple is not used, then a separate mea-
surement of the combustion-air temperature must be made
using the probes thermocouple during the analyzer’s warm-
up period.
Inspect the flue-gas hose for cracks. Replace the hose if it is defective.
Before using the analyzer, check that the Water Trap / Filter is dry and
not dirty. If necessary, dry out the trap and replace the filter element per
Section 5.8.