Specifications
Instruction 24-9388
4
Fyrite Pro
Power Requirements ........................ Four disposable ‘AA’ Alkaline bat-
teries (an optional AC adapter is
available that eliminates the need
for batteries)
Operating Time (using batteries) ... Up to 18 hours continuous (pump
running and backlight off)
Warm Up Time ................................... 60 seconds
Memory ............................................... Up to 10 sets of readings can be
saved in memory. When memory
is full, the next reading saved will
overwrite the oldest reading.
Display ................................................. 4 Digit, 2 Line, 7-segment Liquid
Crystal Display
Front Panel Controls ....................... Six push button switches
(Refer to Section 3.4)
Accuracy:
• Oxygen............................................. ±0.3% O
2
with a typical flue gas
concentration of CO
2
• Carbon Monoxide ............................ ±5% of reading or ±10 ppm, which-
ever is greater*
• Flue Gas Temperature .................... ±2 °C between 0 & 124 °C
(±4 °F between 32 & 255 °F)
±3 °C between 125 & 249 °C
(±6 °F between 256 & 480 °F)
±4 °C between 250 & 400 °C
(±8 °F between 481 & 752 °F)
• Second Thermocouple Channel ...... ±2 °C between 0 & 100 °C
(±4 °F between 32 & 212 °F)
• Ambient Temperature..................... ±2 °C between 0 & 40 °C
(±4 °F between 32 & 104 °F)
• Draft ................................................ ±1% of reading or ±0.05 mB
(±0.02 inWC), whichever is greater
* Tighter CO accuracy in the lower ranges, up to ±2 ppm, may be
attained if a lower range calibration gas (e.g. 100 ppm CO) is used.