Specifications

Instruction 0024-9460
53
Fyrite
®
INSIGHT
Main Diagnostics - Lists the current status of the Stack and Air Ther-
mocouple channels, Reference Temperature channel, Pressure channel,
CO channel, O
2
channel and Battery.
O2 Sensor Life - Displays the approximate remaining life of the Oxygen
sensor.
Fresh Air Diagnostics - Fresh air diagnostics will cycle the instrument
through the 60 second warm-up to check on the status of the sensors. The
probe must be in fresh air to perform a valid check.
Access the DIAGNOSTIC menu as follows:
1. Display the MAIN MENU by pressing the MENU (F2) button. If neces-
sary, press ESC until MENU appears above F2.
2. Use the ▲▼ buttons to highlight DIAGNOSTICS and then press ENT to
display the DIAGNOSTIC MENU.
3. Use the ▲▼ buttons to highlight the desired diagnostic topic and then
press ENT to display the information under that topic.
4. Press the ESC key to exit back to the DIAGNOSTIC MENU or the
MENU (F2) key to exit back to the MAIN MENU.
6.10 Status Screen
The Status Screen provides a quick reference to key items when trouble-
shooting:
• Version
• ADC Version
• Boot Version
• Model
• SN
Access the Status menu as follows:
1. Display the MAIN MENU by pressing the MENU (F2) button. If neces-
sary, press ESC until MENU appears above F2.
2. Use the ▲▼ buttons to highlight STATUS and then press ENT to dis-
play the DEVICE STATUS screen.
3. Press the ESC key to exit back to the MAIN MENU.