Specifications

Instruction 0024-9460
50
Fyrite
®
INSIGHT
7. Strip
1
/4 inch of insulation from each of the probe’s thermocouple
connector wires.
IMPORTANT: In Step 8, the thermocouple wires must first be twisted
together and then crimped.
8. Twist both red thermocouple wires together; insert them into the
supplied wire-splice connector; and then crimp the connector using
a pair of pliers. Repeat this step for the yellow thermocouple wires.
9. Reassemble the probe handle, being careful not to pinch the ther-
mocouple wires between the handle pieces. In addition, ensure that
the end of the thermocouple is in front of the rib molded into the
bottom handle piece; otherwise, the handle pieces will not fit tightly
together.
6.6 Gas Pump Replacement
Parts & Tools Required:
Replacement pump (P/N 24-3049)
Medium phillips screwdriver
Procedure:
1. Gain access to the gas pump by removing the rear case. Refer to Sec-
tion 6.1.
2. Disconnect J9 (pump) and J11 (Battery) connectors.
3. Lift the circuit board out of the unit.
4. Remove the screw that secures the pump to the board in order to
separate the pump from the board.
5. Remove tubing from pump.
6. Install new pump by reversing this procedure.
6.7 Cleaning the Probe
The probe tube and gas-sample hose will become dirty under normal use.
Note that the water trap’s filter element should prevent soot from reaching
the analyzers internal components. If the probe is not kept clean, it could
become clogged and restrict the flow of gas into the analyzer, resulting in
incorrect combustion test readings and calculations.