If the sensor’s output is too low to be usable, then the message
Bad Calibration Sensor End of Life, Entry Not Saved” will
appear. The sensor will now be marked as being BAD in the
DIAGNOSTICS screen.
6. Power OFF the regulator of calibration fixture and remove the CO
cylinder.
6.0 Maintenance
6.1 Fyrite
®
INSIGHT Disassembly
The following section describes how to disassemble the Fyrite INSIGHT to
perform necessary periodic maintenance. (See Figure 6.1 for diagram.)
Tools Required:
• Medium Phillips screwdriver
Procedure:
1. Unplug all thermocouples from bottom of analyzer.
2. Remove battery cover and then remove batteries.
3. Remove sensor caps, disconnect tubing, and then unplug all sensors.
4. Lay analyzer face down on a padded work surface; and then using
a medium Phillips screwdriver, remove the unit’s four rear-case
screws.
5. Lift rear case from analyzer and set aside.
6. Unplug electrical connector J11 from printed circuit board.
7. Lift printed circuit board from analyzer.