5.7 CO Sensor Calibration
Material required:
• Calibration kit, P/N 0024-7059
• Gas cylinder: 500 ppm CO in air, P/N 0024-0492
To improve the accuracy of the CO reading, we suggest that if the analyzer
will be primarily used for flue gas testing, then calibrate using 500 ppm
CO. If the analyzer, however, will be pri ma ri ly used for ambient testing,
then calibrate using 100 ppm CO.
Procedure:
1. If not already done, turn ON the analyzer and
display the CALIBRATION LIST screen per
Section 5.2.
2. Use the
buttons to highlight CO, and
then press ENTER to display the CALI-
BRATE CO screen.
“Measured” is the current CO reading,
while “Applied” is a known CO level
that will be applied for calibration
purposes.
3. Attach a 500 ppm CO cylinder to the regula-
tor of the calibration fixture.
4. Use the
and buttons to enter an Applied value that exactly
equals the concentration stamped on the CO cylinder.
The calibration range is from 20 to 1,000 ppm. An attempt to
calibrate outside this range will cause the message “Applied
Value High” (or Low) to appear at the bottom of the screen.
5. Wait until the Measured reading stabilizes and then press ENTER
to calibrate the CO Measured value to that of the Applied value.
The message “Good Calibration” should briefly appear.
If the sensor’s output is low, but still usable, then the message
“Good Calibration WARNING Low Sensor” will appear. The
sensor will now be marked as being Low in the DIAGNOS-
TICS screen.