Specifications
Instruction 0024-9460
41
Fyrite
®
INSIGHT
The calibration range is from 175 to 625 °F (80 to 330 °C). An at-
tempt to calibrate outside this range will cause the message “Applied
Value High” (or Low) to appear at the bottom of the screen.
7. Wait until the Measured reading stabilizes, and then press ENTER to
calibrate the TS-Span Measured value to that of the Applied value;
after which the message “Good Calibration” should briefly appear fol-
lowed by the CALIBRATION LIST screen being re-displayed.
5.6 T-Air Calibration
This procedure first zeros and then spans the ambient-temperature chan-
nel to known temperature values.
The use of an electronic thermocouple simulator is the preferred method of
producing the desired calibration temperatures. Alternatively, ice and boil-
ing water baths can be used.
Material Required:
• Thermocouple simulator (K-type)
- Range: 0 to 600 °F
- Accuracy: ±0.5 °F
• (Alternatively) Ice-water, boiling water, thermometer
TA-Zero Procedure:
1. Set thermocouple simulator to room temperature and plug its output
into the T-AIR connector located at the bottom of the analyzer.
Alternatively: Plug the probe’s thermocouple
into the T-AIR connector located at the bot-
tom of the analyzer. DO NOT attach the
probe’s gas hose to the analyzer’s GAS
port; otherwise water will be drawn
into the analyzer!
2. If not already done, turn ON the analyzer
and display the CALIBRATION LIST
screen per Section 5.2.
3. Use the buttons to highlight T-Air, and
then press ENTER to display the CALI-
BRATE TA-ZERO screen.