2. If not already done, turn ON the analyzer and display the CALI-
BRATION LIST screen per Section 5.2.
3. Use the ▲▼ buttons to highlight T-Stack,
and then press ENTER to display the CALI-
BRATE TS-ZERO screen.
"Measured" is the current temperature read-
ing, while "Applied" is a known temperature
that will be applied for calibration purposes.
4. Set thermocouple simulator to 32 °F (0 °C),
and then use the
and buttons to
enter an Applied value that exactly equals
the setting of the simulator.
Alternatively: Submerge probe tip into an ice-water bath with a ther-
mometer, wait several minutes, and then use the and buttons
to enter an Applied value that exactly equals the thermometer read-
ing.
The calibration range is from 32 to 41 °F (0 to 5 °C). An attempt to
calibrate outside this range will cause the message “Applied Value
High" (or Low) to appear at the bottom of the screen.
5. Wait until the Measured reading stabilizes,
and then press ENTER to calibrate the
TS-Zero Measured value to that of the Ap-
plied value; after which the message “Good
Calibration” should briefly appear followed
by the CALIBRATE TS-SPAN screen.
TS-Span Procedure:
6. Set thermocouple simulator to 572 °F (300 °C), and then use the
and buttons to enter an Applied value that exactly equals the
setting of the simulator.
Alternatively: Submerge probe tip into a container of boiling water
with a thermometer, wait several minutes, and then use the and
buttons to enter an Applied value that exactly equals the ther-
mometer reading.