5.3 B-Smart Sensor Replacement & Calibration
To replace a B-Smart sensor and calibrate do the fol-
lowing:
1. Enter the CALIBRATION MENU per Sec-
tion 5.2.
2. Use the ▲▼ buttons to select B-SMART.
Press ENTER to display the B-Smart code
screen.
3. Use the ▲▼ buttons to enter the 10 digit al-
phanumeric code supplied with the B-Smart
sensor. Use the the ◄► buttons to move the
cursor across the screen. Press ENTER.
NOTE: If the correct code was entered, the analyzer accepts it and
returns to the CALIBRATION MENU. If an incorrect
code was entered, the screen will display "Invalid Code".
Check to make sure the correct code has been entered. If
problem persists, contact your nearest Bacharach Service
Provider.
Bacharach also offers a convenient Exchange Program that allows the cus-
tomer to regularly send in old sensors and, as scheduled, receive new re-
placements that are already calibrated and that include a code that can be
entered into the analyzer for a quick convenient setup. Contact Bacharach
customer service for more details about this program.
5.4 Pressure Sensor Calibration
This procedure calibrates the pressure sensor to a known pressure value.
Materials Required:
• Bellows
• Manometer
- Range: ± 8 in. of water column (± 20 mb)
- Accuracy: ± 0.01 in. of water column (± 0.025 mb)