Specifications
Instruction 0024-9460
36
Fyrite
®
INSIGHT
2. Allow the analyzer to cycle through its 60 second warm-up period.
During warm-up the analyzer's operation is checked and the sensors
are set to the following ambient conditions:
• Oxygen sensor spanned to 20.9%
• CO sensor zeroed
• Pressure sensor zeroed
Any errors detected during warm-up will be listed on the dis-
play immediately following warm-up. Correct any errors before
proceeding. Refer to Section 6.8 for a list of error messages and
their meaning.
3. Display the MAIN MENU by pressing the MENU (F2) button. If
necessary, press ESC until MENU appears above F2.
4. Use the ▲▼ buttons to select CALIBRA-
TION. Press ENTER to display the Calibra-
tion Password screen.
5. Before calibration can begin a 4-place nu-
meric password must be entered. Use the
▲▼ and ◄► buttons to enter the password.
(Note that the default password is 1111)
6. Press the ENTER button to accept the
password. If the correct password was
entered the CALIBRATION Menu will be
displayed.
7. Use the ▲▼ buttons to select the desired sensor channel to be
calibrated, and then perform the calibration procedure for that
sensor as described in the following sections.