Specifications

Instruction 0024-9460
15
Fyrite
®
INSIGHT
6. Repeat Steps 3-5 to edit the remaining lines.
7. Once the input is complete, use the ▲▼ buttons to select EDIT
COMPLETE.
8. Press ENTER to accept the entered information or ESC to revert to
the previous information and display the SETUP MENU.
Helpful Hint: Use the Fyrite
®
user software to upload this informa-
tion from your computer.
To Clear a Username proceed as follows:
1. Enter the SETUP MENU per Section 3.5.1. If necessary, press ESC
until MENU appears above F2.
2. Use the ▲▼ buttons to select USERNAME. Press ENTER to dis-
play the most recently saved information.
3. Clear the information by selecting the CLEAR option (F3).
3.5.9 Run/Hold Format
Fyrite INSIGHT test data is located in the Run/Hold screen. By press-
ing the RUN/HOLD key, you should hear the pump running and see the
word RUN at the upper-left hand corner of the display. The instrument
is continuously measuring and calculating the data that is shown in the
Run/Hold screen. Press the RUN/HOLD key again. The pump should stop
running and the word HOLD should be shown at the upper-left hand cor-
ner of the display. The instrument will now show the last measured and
calculated data taken before the instrument was placed in HOLD. Use the
▲▼ buttons to scroll through the complete list of measured and calculated
values when the instrument is running or in the HOLD mode.
The order in which data appears in the RUN/HOLD screen is as follows:
O
2
= Oxygen
CO = Carbon Monoxide
Eff = Combustion Ef ciency
CO
2
= Carbon Dioxide
T-STK = Stack Temperature
T-AIR = Ambient/Primary Air Temperature
EA = Excess Air
CO(n) = Carbon Monoxide content referenced to an Oxygen
percentage
NOTE: (n) is the current O
2
Reference selected.