Specifications

Instruction 0024-9460
13
Fyrite
®
INSIGHT
3.5.5 O
2
Reference
The measured value of CO can be referenced to a specific O2 percentage of
0% to 15%. Set up the reference value as follows:
1. Enter the Setup Menu per Section 3.5.1. If
necessary, press ESC until MENU appears
above F2.
2. From the SETUP MENU, use the ▲▼
buttons to select O2 REFERENCE. Press
ENTER to display the O2 Reference Screen.
3. Use the ▲▼ buttons to enter the O2 refer-
ence value.
4. Press ENTER to save the selection and re-
turn to the SETUP MENU or ESC to revert
to the previous setting.
3.5.6 Print Pressure
Select whether to print or not print the pressure measurement on the com-
bustion test printout as follows:
1. Enter the Setup Menu per Section 3.5.1. If necessary, press ESC
until MENU appears above F2.
2. From the SETUP MENU, use the ▲▼ but-
tons to select PRINT PRESSURE. Press
ENTER to display the O2 Reference Screen.
3. Use the ▲▼ buttons to select YES or NO to
print the pressure measurement.
4. Press ENTER to save the selection and re-
turn to the SETUP MENU or ESC to revert
to the previous setting.
3.5.7 Zoom
Combustion test data in the Run/Hold screen can be shown with enlarged
characters to make viewing easier. The operator can set zoom levels to
Standard, 2X or 3X. The Standard zoom setting will display 4 lines of
combustion test data at one time; 2X will display 3 lines of data with en-
larged characters; 3X will display 2 lines of data with enlarged characters.
The operator can scroll through the complete list of measured and calcu-