Specifications

Monoxor
Plus Manual Troubleshooting
0019-9370 Rev 0 48
Section 6. Troubleshooting
6.1. Error and Warning Messages
Message Description
Low Sensor CO CO sensor output was low but still usable. Sensor may need to be
replaced in the near future.
Low Battery Battery voltage is low. Replace the batteries.
Applied Value
High/Low
An attempt was made to calibrate a sensor outside its range
either above (High) or below (Low) the acceptable range.
Warm-up
Sensor Error
CO sensor was not zeroed at warm-up because of high output.
Run instrument on fresh air then restart instrument to re-zero
sensor. If the message persists, the CO sensor may need to be
replaced.
Temperature sensors are measuring temperature outside the
range of -4° to 212° F at warm-up. Make sure that the
thermocouple is sampling ambient room air within the
temperature range at warm-up.
Messages will indicate which sensors are in error.
Set Clock Time and date values need to be set in the instrument.
X X X Occurs in the number fields of sensors that have achieved over-
range condition.
* * * Occurs in the number fields of sensors. Replaces in-error sensor
values and any calculated values that depend on those sensor
values.
- - - Occurs in the number fields of sensors and indicates that values
were not calculated.
Beeping Slow beep frequency: Replace batteries warning.
Fast beep frequency: CO alarm is active and set.
NOTE: If a particular sensor is in error during warm-up, the
instrument automatically displays the error. The instrument
continues to operate with the sensor in error, however
information dependent on the sensor in error is not displayed.